Fluke 37X Clamp Meter recall
The printed circuit assembly may not be properly fastened to the test lead input jack. This may result in inaccurate voltage readings, including a low or no voltage reading on a circuit energised with a hazardous voltage, presenting a shock, electrocution or thermal burn hazard.
Immediately stop using your affected 37X series clamp meter(s) even if you have not experienced problems.
ADVERTISEMENT
To determine if your clamp meter is affected by this recall notice, visit this link, enter your model and serial number in the box at the bottom of the page and click “Check.”
Any serial number preceded or followed by the letter “R” is not affected by this recall notice.
*If a serial number is not available, your product is part of the recall. Follow the recall instructions below.
If you own one of these clamp meters, please stop using it and send it back to Fluke for repair. If you are not the primary user of the Fluke 37X Series Clamp Meter, please pass this notice along to the appropriate people within your organisation.
(Please only send the clamp meter – any test leads/boxes/manuals that are sent with the unit will NOT be returned.)
-